ScanEM Probes
Specification

Parameter

ScanEM-C Probes
CTK015

ScanEM Probes
CTK010

 

The main difference between ScanEM-C and ScanEM probes is that ScanEM probes do not provide AC/DC output and work only in a self-contained mode as E- and H-field "sniffers."

Probe

ScanEM-EC

ScanEM-HC

ScanEM-E

ScanEM-H

Field

Electric

Magnetic

Electric

Magnetic

Frequency Range

2MHz..2GHz

1MHz..1GHz

2MHz..2GHz

0.1..100MHz

Self-Indication

5-LED color level bar; Speaker (tone pitch proportional to the field strength)

Output

AC and DC; SMB jack

N/A

Sensitivity

-10dBm/(V/m)

-10dBm/mT

 

AC Output Range

0..1V p-p

DC Output Range (>1MOhm load)

0..2V

Dimension (approx.)

5.75" x 1.21" x 0.76" (146mm x 31mm x 20mm)

Weight (approx.)

2oz (57g) each

Batteries

2 AAA (included)

Content of ScanEM Kits (see separate page for ScanEM-Q)

ScanEM-C kit model CTK015

ScanEM kit model CTK010

  • ScanEM-EC probe
    ScanEM-HC probe
    6' (1.8m) SMB/BNC cable
    Heavy-duty plastic storage case
    User's Guide
    Batteries
  • ScanEM-E probe
    ScanEM-H probe
    Heavy-duty plastic storage case
    User's Guide
    Batteries
  •  

    ScanEM-EC and HC Frequency Response vs. "Standard" Passive Probes

    Ó 1999 Credence Technologies, Inc. Mail comments to webmaster@credencetech.com
    Aware, EMAlert, EM Event, EMEye and ScanEM are trademarks of Credence Technologies, Inc.
    Products are covered by patents or are patent pending. All specifications are subject to change without notice.